TW-72Do not attempt to use or maintain thisunit until you read and understand theseinstructions. Do not permit untrainedpersons to use or maintai
Maintenance Read the Safety Precautions in the front of this manual before attempting any repairson these containers. Also, follow these additional sa
3. Allow the container to stabilize for 24 hours, then weigh it. Record the weight, time and date.4. Reweigh after the recommended 48 hours. The t
FULL VIEW CONTENTS GAUGE MAINTENANCEThe contents of these containers is measured with the Full View Contents Gauge. Thedevice consists of the gauge as
If the rod is turned clockwise (to the left) with respect to calibration nut, the exposedportion of rod becomes longer and the gauge yellow band will
Lower the assembly about 4 in. (102 mm) slowly and try to keep the rod in the center ofthe threaded entrance hole as you do. If you are careful during
TroubleshootingSymptom Possible Cause Corrective ActionConsistently low operatingpressure.1. Relief valve open at lowpressure.2. C o ld liqu id .1.
Replacement PartsAccessoriesAccessories available for use with Taylor-Wharton XL Series containers include:· Manifolds, Automatic and Manual· Transfer
IndexNo.DescriptionPart No. Recommended for 10Units1. Gasket, Glass Filled Teflon 7701-0083 5 each2. Float, Contents Gauge 1604-9C60 1 each3. Conten
TAYLOR-WHARTONXL Series Liquid CylindersTaylor-Wharton Gas Equipment Division, Harsco Corporation warrants thateach of its Refrigerator and Dewars wil
Container Safety ... 1General Information ...
Pressure Hazard – The containers covered by this literature contain liquefied gasunder pressure. Sudden release of this pressure may cause personal in
The XL-160 and XL-180 are vacuum insulated, stainless steel containers designed tostore, transport, and dispense cryogenic liquid nitrogen. Built to D
Handling the ContainerXL Series containers are very rugged liquid cylinders. All cryogenic liquid containershave an inner container and on outer conta
OperationThe model numbers of XL-160/XL-180 indicate their respective liquid storage capacitiesin liters of product. Both cylinders are designed for l
Withdrawing Liquid From The ContainerTo use the container in liquid delivery service, attach a transfer hose to the LIQUIDconnection and open the adja
Filling the Container by Volume1. Visually inspect the container. Do not attempt to fill containers that have broken ormissing components.2. Connect a
In use, the CGA end fitting on the Fill/Vent Tee Assembly couples to the fill connectionto be filled. The Relief Valve vents pressure over 350 psig (2